YSOT-LED5070-R9 Medical equipment LED shadowless operation lamp
Main specifications Diameter of Light Head (mm): 700 and 500; Quantity of LEDS: 80 and 48; Illumination (10 Steps Adjustable): 40~160klx and 40~160klx; Color rendering index: Ra>95 and Ra>95; Color temperature (5 Steps Adjustable): 3500-5000K and 3500-5000K
1. Expected use This product is used in various surgical lighting in hospital operating room.It can also be use in medical lighting situations in emergency room.
Conditions of use
2.1 Normal working conditions Temperature:10ºC~30ºC; Relative humidity:≤70%; Atmospheric pressure range:860hPa~1060hPa;
This equipment can not be worked in flammable and explosive gas environment. The type of device security: I type equipment; Working system: continuous operation; Liquid injection protection level: IPXO.
2.2 Power parameters Power supply voltage:AC220(±10%)V; Frequency:50Hz; Input power:150VA; PLM-700 main lamp power:80W; PLM-500 main lamp power:70W.
2.3 suitable mounting height:3200~3300 mm Low-storey may bring inconvenience to users. please refer to the product diagram. in special circumstances, you should contact with the manufacturer when purchasing.