Type: | Anaesthesia Machine |
---|---|
Kind: | Surgery |
Certification: | CE |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Technical Specification
• Dimensions(H×W×D): 1403×704×750 mm
(without YOKE)
• Weight: 110kg (basic unit)
• Operation Temperature: 10~40°C
• Operation Humidity: 80% (noncondensing)
• Storage Temperature: -20~60°C
• Storage Humidity: 95% (noncondensing
Electrical Specifications
• Mains: AC 100~240V, 50/60Hz
• Battery: DC 24V 4Ah, minimum 120 min
• Outlets: 4 on back, 1.5A individual
Pneumatic Specifications
• Gas Supply: 280 ~ 600kPa
• Electronically controlled mixer
0~18L/min
• Or set each gas independently
O2: 0~10L/min
N2O: 0~10L/min
Air: 0~10L/min
• Auxiliary O2: 0 ~15L/min
• Gas System: Oxygen supply failure protection
• O2 flush: 25~75L/min
• Gas cylinder Optional
• ACGO: Optional auxiliary common gas outlet
• AGSS: Optional
Breathing System Specifications
• Operating Modes: Closed, Semi-closed, Semi-open
• CO2 Canister: single canister 1.5L
• APL Range: Spontaneous breathing (SP) -70 cmH2O
• Material: Autoclavable (except O2 cell and pressure
gauge)
• CO2 bypass: Optional
Ventilator Specifications
Driven Mode: Pneumatically driven and Electronically
controlled
• Patient Types: Adult and Pediatric
• Ventilation Modes: IPPV, PCV, SIMV-VC, PS/CPAP and Manual
optional: SIMV-PC,PCV-VG,SIMV-VG
• Man./Vent. switch: One key switch
• Tidal Volume: 10~1500ml
• Ventilation Frequency: 2~100bpm
• I:E: 4:1~1:8
• P(inspired) Range: 5~70cmH2O
• P(support) Range: 3~50cmH2O
• Inspiratory Time: 0.2 5s
• PEEP: OFF, 3~30 cmH2O
• Tslope: 0~2s
• Inspiratory Plateau: OFF, 5%~60%Ti
• Trigger: 1~15L/min
• Compensation: Compliance and leakage compensation,
fresh gas compensation, altitude compensation
Ventilator monitoring & alarm
• Pressure: Paw, Ppeak, Pplat, Pmean, PEEP
• Volumes: MV, Vt, Freq
• Gas: FiO2, optional: CO2, AG, MAC
• Graph Display: P-t, F-t, V-t waveform
P-V, F-V, P-F loop
• Alarm: Audible and Visual alarms
High/Low Airway Pressure
Continuous airway pressure
High/Low Minute Volume
High/Low FiO2
High Freq.
O2 Supply failure, Power failure
Apnea
• Alarm Silence: 120 seconds
Vaporizer
Support two vaporizers (optional standby vaporizer parking
holder)
• Agent: Sev. Enf. Iso. Halo. Des.
• Installation Mode: Selectatec® with interlock
• Filling Type: Pour-Fill, Key-Fill, Quik-Fil®