Type: | Automated External Defibrillator |
---|---|
Kind: | Automated External Defibrillator |
Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Category
|
Specification
|
Output
|
Biphasic Truncated exponential
|
Energy Sequence(adult )
|
150/150/200(Default)
150/200/200 200/200/200 Note: the Energy sequence can be changed using the computer IR port. |
Analysis Time
|
9 sec.
|
Energy Sequence(Pediatric)
|
50/70/85(Default)
70/70/85 70/85/85 85/85/85 Note: the Energy sequence can be changed using the computer IR port. |
Charge Time
|
10 sec. to 200 J
|
The maximum time from the initiation of rhythm analysis to readiness for maximum Energy discharge after 15 shocks.
|
Less than 20 seconds
|
The maximum time from initially switching power on to readiness for maximum Energy discharge after 15 shocks..
|
Less than 30 seconds
|
Audible Prompts
|
19 audible prompts
|
Visual Prompts
|
LED prompts; Automatic self-test
|
Controls
|
six buttons - On/Off, Shock, and 4 software-configurable buttons
|
Output Energy Accuracy
|
±15% into any impedance from 25 to 175Ω
|
The Maximum Voltage
|
1800±20V
|
Output disabled when the PATIENT impedance is outside limits
|
20Ω to 200Ω
|
Recovery of the ECG input after defibrillation
|
<10s
|
Specificatio