Medical Equipment YSBB-200C Neonatal Transport Infant Incubator
STANDARD FEATURES:
Servo Control Mode: air temperature, skintemperature;
Parameter Displaying: set temperature,air temperature, skin temperature, internal humidity, heat power rate;
Audible & Visual Alarms: powerfailure, sensor failure, higher & lower temp, air & skin temperature override,fan failure, system failure;
>37ºCPreset Mode: power failure reset memory, alarm silence, keypad lock, self-testfunction;
Hood with dual slope; 6 hand-ports;
RS-232 Output Connector (use for networksystem);
IV poles;
Protective guard rail;
Storage cabinets.
Specifications of hood
Mattress to Hood: 450 mm
Observation side of hood: Two incline
Operate port: 6 (2 Iris ports)
Tubing grommets: 4
Front access panel: 826×310 mm
Mattress Tray Size: 630×360 mm
Technical Specifications:
Powerrequirement ≤400VA
Temperaturecontrol mode: Air and skin servo control
AirTemp Control Range: 20ºC~37ºC
Temperatureoverride mode to: 20ºC~39ºC
SkinTemp Control Range: 30ºC~37ºC
Temperatureoverride mode to: 30ºC~38ºC
Skintemp sensor accuracy: ≤0.3ºC
Temperaturefluctuation: ≤0.5ºC
Mattresstemperature uniformity: ≤0.8ºC
Temperaturerise time: ≤45 min
Noiselevel within hood environment: ≤55dB
Alarms
Power failure alarm
Fan failure alarm
Sensor failure alarm
High air temperature alarm: +3.0ºC
Low airtemperature alarm: -3.0ºC
High skin temperature alarm: +1.0ºC
Low skin temperature alarm: -1.0ºC
Over air temperature alarm: ≤38.0ºC; ≤40.0ºC
Over skin temperature alarm: ≤38.0ºC; ≤39.0ºC
Systemfailure alarm
Alarmsilence function
Operational condition
Environment temperature: +20ºC~+30ºC
Environment relative humidity: 5%~99%
Environment air velocity of flow: <0.3m/s
Others
Infant bed tilt adjusting ±10° step-less
Mattress tray Max. load 10kg
Transfusion pole Max. load 2kg
CabinetMax. load 6kg