Certification: | CE |
---|---|
Scale: | Medium |
Type: | Infusion Pump |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
PARAMETER |
SPECIFICATION |
Drug library |
10 categories of drugs, amounts to 1030 kinds of drug names, 17kinds of dose unit, such as ug/kg/min, mg/kg/min,etc. Makes drug dispensing more convenient and safer.) |
History record function |
Record infusion start time, infusion over time, infusion rate, alarms, drug name, dose and bolus volume, etc. |
Drop Detector |
Outside caliper rotary detector (patent) Infrared sensor |
Control Way |
Drop counting and adjusting the delivery rate automatically |
Fit IV tubes |
20d/ml,60d/ml and other, as well as brown tubes |
Pump Type |
Peristaltic finger pump |
Accuracy |
±2%, when delivery rate ≤25drop/min (drop counting & time control mode) ±3%, when delivery rate >25drop/min (drop counting & time control mode) ±5% (Volume control mode after the correct calibration) |
Delivery rate range |
0.1ml/h~1200ml/h (0.1ml/h~99.9ml/h, 0.1ml/h step; >99.9ml/h, 1ml/h step) |
Delivery time range |
1min~1999min (1 min step) |
Volume limit range: |
0-9999ml (1 ml step) |
Total Volume Range |
0-9999.9ml (1 ml step) |
Purge Rate |
600.0ml/h |
KVO Rate |
KVO is 1ml/h when the delivery rate≤10ml/h, Otherwise KVO is 3ml/h. |
Occlusion pressure |
Three level setting LOW, MEDIUM or HIGH |
Alarms |
Near empty, Syringe empty, Volume complete (when limit volume is set), Occlusion, Syringe loose, Operation error, Low battery, device failures. Three levels of alarm volume LOW, MEDIUM or HIGH |
Bolus function |
Bolus rate:10~1200ml/h (pre-selected),Bolus volume: maximum of 100ml |
Power source |
AC power supply: AC 100 to 240V( +/-10%),50/60Hz; 12 hours of continuous operation |
Accessories: |
AC power line, Operational manual |
Protection Class |
Class I, Type CF, IPX4 |
Dimension /Net weight |
130mm*200mm*230mm , 2.5kg |