Type: | Anaesthesia Machine |
---|---|
Kind: | Surgery |
Certification: | CE |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product Description
• PCV and SIMV+PS modes ensure flexibility in ventilation, suitable for both pediatric and adult operation
• Flowmeter with backlight, suitable for endoscopy surgery etc. • Breathing system warmer for avoid condensation build-up
during low-flow anesthesia
• Enhances usability with auxiliary oxygen, suction, auxiliary
power source, monitor arm, infusion pump etc.
• Optional functions monitoring O2, CO2 and anesthetic gas • With CE certificate, meets EU clinical requirement
Physical & Environment Specifications | |
Dimensions(W×H×D) | 704mm × 1403mm × 750mm |
Weight | 100kg (basic unit) |
Storage Humidity: | ≤93% (noncondensing) |
Operation Humidity | ≤80% (noncondensing) |
Electrical Specifications | |
Mains | 100~240V AC, 50/60Hz |
Battery | 24V DC, minimum 120 min |
Outlets | 3 on back, 1.5A individual |
Pneumatic Specifications | |
Gas Supply | Air, O2, N2O; 280kPa ~ 600kPa |
Flowmeters | O2 Range: 0~1L/min; 1~10L/min N2O Range: 0~1L/min; 1~10L/min Air Range: 0~1L/min; 1~10L/min Auxiliary O2 Range: 0 ~12 L/min |
Gas System | Flowmeter backlight Oxygen supply failure alarm N2O-cut-off system O2 flush: 25~75L/min |
ACGO | Optional auxiliary common gas outlet |
AGSS | Optional anesthetic gas scavenging system |
Breathing System Specifications | |
Operational Modes | Closed, Semi-closed, Semi-open |
CO2 Canister | 2 Loose fill (3L) or 2 prepacks, optional single canister (1.8 L) |
APL Range | Spontaneous breathing (SP) -70 cmH2O |
Material | Autoclavable (except O2 cell and mechanical pressure gauge) |
CO2 bypas | Optional |
Ventilator Specifications | |
Driven Mode | Pneumatically driven and electronically controlled |
Patient Types | Adult, Pediatric |
Ventilation Modes | IPPV, PCV, SIMV+PS, PS and Manual |
Setting Modes | Touch key & navigator wheel knob |
Tidal Volume | 20~1500ml |
Ventilation Frequency | 4~100bpm |
I:E Ratio | 4:1~1:8 |
Pressure (inspired) Range | 5~70cmH2O |
Pressure (support) Range | 3~50cmH2O |
PEEP | Off,3~30 cmH2O |
Tslope | 0~2s |
Inspiratory Plateau | Off, 5%~60%Ti |
Trigger Sensitivity | 1~15L/min |
Compensation | Compliance and leakage compensation, fresh gas compensation and altitude compensation |
Ventilator Monitoring | |
Display | 10.4" TFT color screen |
Pressure | Ppeak, Pplat,Pmean,PEEP |
Volumes | MV, Vt, Freq |